BSL laboratory
FUME HOOD SUPERSTRUCTURE
Fume Hood Inner Panels & Baffles 6mm thick panels shall be made of laboratory grade Phenolic Resin / Compact Laminate to have property of excellent chemical & heat resistance and fire retardant. These panels shall be in rigid construction with smooth finish and easy clean ability.
BASE CABINET
Twin cabinets in Zero Spangled Galvanized Iron (GI) construction for apparatus storage cabinet. Cabinets are mounted on a common base frame with castor wheels. Each cabinet to have twin door in double skin construction. Each cabinet shall have one adjustable
ACCESSORIES
Airflow Monitor This Face Air Velocity monitor gives digital display of face velocity in m/sec or fpm with on screen display for Safe and Alarm conditions with audible alarm and LED indication.It has a feature of Push‐button calibration and configuration along with Plug‐in connections .
LABORATORY FUME HOODS
We manufactures specialty and custom fume hoods for the safe handling of chemical fumes and volatile compounds for a range of laboratory, cleanroom and industrial applications. Our units are designed for your process requirements taking into account your facility and existing HVAC system. We also manufacture spot/source-capture filtered cleanroom models and Class I, Division 1 explosion-proof electrical components.
- Highly energy efficient direct exhaust fans
- Effective exhaust efficiency
- Structural design
- Flip on type front foil in SS construction
- High standard service fittings
- Full length sash handle
- Unique design sash movement mechanism
- Elegant looks, quiet & ergonomic
- Bench, low bench and walk in hoods
Comparison statement for salient features of fume hood with other manufacturers
FUME HOODS |
||
MATERIAL
DESCRIPTION |
OTHER MANUFACTURERS | Labnation India Pvt Ltd. |
Base Raw
Material |
Mild Steel (MS) or CRCA material,
poor corrosion resistance |
Zero Spangled Galvanized Iron (GI). Better corrosion resistance than any MS material |
Airfoil | CRCA MS powder coated, starts
corroding in very short time due to daily wear and tear process |
2mm thick Stainless Steel (SS). This part is more prone to wear and tear, hence SS only |
Airfoil Design | Not provided by many manufacturers | Flip‐open type design for ease of maintenance and better cable management |
Cable Manager | Not provided by many manufacturers
|
PVC cable manager provided to avoid any cable dangling from equipment to sockets |
Service fittings | Not manufactured in house, outsourced from other vendors, hence
no quality control. Tubing are not included for final connections to the headers. |
In‐House manufactured and are provided with 1.5m long tubing in PU4 and SS Breaded Teflon hose as per service requirement for final connections to headers |
Base Cabinets | Fixed type not convenient for
maintenance, any maintenance has to be done from behind and stopping the working fume hood. |
Provided with Castors, Easily removable type for easy maintenance without affecting the working of the fume hoods. |
VFD for blowers | Not provided, motor runs only at
predetermined speed. Safety of the end‐user is compromised |
Provided for adjusting motor speed to obtain safe exhaust CFM and Air velocity. Safety of the end‐user is not compromised |
Impellers | Local manufactured, very noisy and | Imported, Direct driven, very silent and efficient |
Special Feature |
Not provided by many |
Provision for Co2, Dry powder based fire extinguisher with heat sensor bulbs |
Storage Cabinets
I am text block. Click edit button to change this text. Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.